Keywords: EKG calibration, adjust speed and amplitude, electrocardiogram settings, EKG interpretation, cardiac diagnostics
Before delving into the adjustments, it is essential to grasp the standard settings. These standardized settings are universally recognized and offer a consistent framework for interpreting EKGs across various machines and clinical environments. Typically, EKG machines generate calibration markers, also referred to as standard calibration pulses or signals, either at the beginning or the end of the strip. These markers serve as visual references for both speed and amplitude.
Standard Calibration Pulse
• Appearance: A rectangle or square wave that is 10 mm tall (vertical) and 5 mm wide (horizontal).
• Indicates:
• Amplitude: 10 mm/mV (since 10 mm equals 1 mV)
• Speed: 25 mm/second (since 5 mm horizontally represents 0.20 seconds)
These setting may also be printed on the sheet.
Recognizing Calibration Variants
Speed Variations
a. Increased Speed (50 mm/second)
Calibration Pulse:
• Width Doubles: The marker is 10 mm tall and 10 mm wide.
• Waveform Appearance:
• Stretched Horizontally: Waveforms appear wider.
• Intervals Appear Longer: Time intervals seem extended.
b. Decreased Speed (12.5 mm/second)
Calibration Marker:
• Width Halves: The marker is 10 mm tall and 2.5 mm wide.
• Waveform Appearance:
• Compressed Horizontally: Waveforms appear narrower.
• Intervals Appear Shorter: Time intervals seem shortened.
Amplitude Variations
a. Increased Amplitude (20 mm/mV)
Calibration Marker:
• Height Doubles: The marker is 20 mm tall and 5 mm wide.
• Waveform Appearance:
• Stretched Vertically: Waveforms are taller.
• Voltage Appears Higher: Amplitudes seem increased.
b. Decreased Amplitude (5 mm/mV)
Calibration Marker:
• Height Halves: The marker is 5 mm tall and 5 mm wide.
• Waveform Appearance:
• Compressed Vertically: Waveforms are shorter.
• Voltage Appears Lower: Amplitudes seem decreased.
Quick Tips to Recognize Calibration Variants
Examine Calibration Markers First
• Location: Usually at the beginning or end of each lead or rhythm strip.
• Shape and Size: Changes in the marker’s dimensions directly indicate calibration adjustments.
Look for Printed Calibration Information
• Margins and Headers: Calibration settings are often printed in these areas.
• Common Notations:
• Speed: “25 mm/s”, “50 mm/s”, “12.5 mm/s”
• Amplitude: “10 mm/mV”, “20 mm/mV”, “5 mm/mV”
Analyze Waveform Characteristics
Unusual Waveform Size:
• Too Wide: Suggests increased speed.
• Too Narrow: Suggests decreased speed.
• Too Tall: Indicates increased amplitude.
• Too Short: Indicates decreased amplitude.
When and Why to Adjust EKG Speed
Analyzing Fast Heart Rates (Tachycardia)
Why Adjust?
• Improve Waveform Clarity: In cases of tachycardia (heart rates above 100 bpm), the EKG waveforms can appear compressed, making it difficult to distinguish individual components like the P wave, QRS complex, and T wave.
How to Adjust?
• Increase Speed to 50 mm/second: Doubling the speed stretches out the waveforms horizontally, allowing for better visualization of each cardiac cycle.
At 50 mm/second:
• 1 small square = 0.02 seconds
• 1 large square = 0.10 seconds
How we calculate the heart rate changes accordingly also: Heart Rate = 3000 / Number of small squares between R-R intervals.
Clinical Benefits:
• Detailed Rhythm Analysis: Enhanced clarity aids in identifying arrhythmias, conduction blocks, or other abnormalities that might be obscured at standard speed.
Evaluating Slow Heart Rates (Bradycardia)
Why Adjust?
• Conserve Paper and View More Cycles: In bradycardia (heart rates below 60 bpm), the waveforms are spread out over a longer period, which can make it challenging to view multiple cardiac cycles on one page.
How to Adjust?
• Decrease Speed to 12.5 mm/second: Halving the speed compresses the waveforms horizontally, allowing more cycles to fit on a single strip.
At 12.5 mm/second:
• 1 small square = 0.08 seconds
• 1 large square = 0.40 seconds
Accordingly, Heart Rate = 750 / Number of small squares between R-R intervals.
Clinical Benefits:
• Long-Term Rhythm Observation: Useful for observing patterns over time, such as in patients with suspected intermittent heart block.
Impact of Speed Adjustment on EKG Interpretation
Understanding Time Intervals
• Standard Speed (25 mm/second):
• 1 small square = 0.04 seconds
• 1 large square (5 small squares) = 0.20 seconds
When and Why to Adjust EKG Amplitude (Voltage)
Detecting Low Voltage QRS Complexes
Why Adjust?
• Enhance Small Waveforms: In some patients, especially those with obesity, pericardial effusion, or pulmonary emphysema, the voltage of the QRS complexes may be abnormally low.
How to Adjust?
• Increase Amplitude to 20 mm/mV: Doubling the amplitude vertically enlarges the waveforms, making small deflections more discernible.
At 20 mm/mV:
• 1 mm = 0.05 mV
• Waveforms appear twice as tall
Clinical Benefits:
• Improved Detection: Enhances the ability to detect and measure small voltage changes, aiding in accurate diagnosis.
Preventing Waveform Overlap in High Voltage Situations
Why Adjust?
• Avoid Clipping of Waveforms: In cases of hypertrophy or abnormal conduction, the QRS complexes may have very high voltage, causing them to overlap or go off the EKG paper.
How to Adjust?
• Decrease Amplitude to 5 mm/mV: Halving the amplitude compresses the waveforms vertically, ensuring the entire waveform fits within the recording.
At 5 mm/mV:
• 1 mm = 0.2 mV
• Waveforms appear half as tall
Clinical Benefits:
• Accurate Measurement: Prevents distortion of waveforms, allowing for precise measurement of amplitudes and intervals.
Impact of Amplitude Adjustment on EKG Interpretation
Understanding Voltage Measurements
• Standard Amplitude (10 mm/mV):
• 1 mm (small square vertical) = 0.1 mV
Practical Considerations
Always Document Calibration Changes
• Prevent Misinterpretation: Failure to note adjustments can lead to incorrect diagnoses due to miscalculations of heart rate and waveform measurements.
Understand the Clinical Context
• Tailor Settings to Patient Needs: Adjustments should be made based on the specific clinical scenario and the patient’s condition.
Recalibrate Formulas and Measurements
• Adapt Calculations Accordingly: Be mindful that standard formulas and normal values apply to standard calibration settings.