Cardiac pacing is a pivotal intervention in cardiology that involves providing electrical impulses to prompt the heart’s activity when its natural electrical system is unable to sustain an adequate rate or rhythm. Pacemakers, the devices that fulfill this role, are indispensable for patients whose conditions hinder the heart’s capability to beat efficiently on its own. For healthcare professionals overseeing the care of these patients, a thorough understanding of pacing mechanisms and the ability to interpret EKG (electrocardiogram) changes are vital.
The Purpose of Cardiac Pacing
The main goal of cardiac pacing is to ensure the heart maintains a proper rate and rhythm to facilitate efficient blood circulation. When the heart’s natural electrical system is impaired due to issues like bradycardia (an abnormally slow heart rate), heart block, or other rhythm disorders, a pacemaker can step in to deliver the necessary electrical impulses. This intervention helps prevent symptoms associated with insufficient heart rates, such as dizziness, fatigue, and fainting, while also lowering the risk of more serious complications.
Mechanisms of Pacing and EKG Changes
Pacemakers function by delivering electrical impulses to specific chambers of the heart, initiating depolarization and subsequent contraction. The mechanism of pacing and its effect on the heart’s electrical activity depend on which chamber is being stimulated and the patient’s underlying cardiac condition. These interventions result in characteristic changes on an EKG, a tool that records the electrical activity of the heart.
When a pacemaker stimulates the heart, it can alter the normal conduction pathways. For instance, ventricular pacing often leads to a widened QRS complex on the EKG due to the delayed spread of the electrical impulse through the ventricular muscle. Understanding these changes is essential for interpreting EKGs in patients with pacemakers and for identifying any potential issues with the device or the heart itself.
Types of Cardiac Pacing
1. Atrial Pacing
Atrial pacing involves delivering electrical impulses to the atria, the heart’s upper chambers, when the natural atrial activity is insufficient. This type of pacing is beneficial for patients with sinoatrial node dysfunction, where the heart’s natural pacemaker fails to generate impulses at an adequate rate.
EKG Interpretation: On an EKG, atrial pacing is indicated by a pacing spike preceding the P-wave, which represents atrial depolarization. The presence of this spike followed by a P-wave shows that the atria have responded to the pacemaker’s stimulus. The subsequent conduction through the atrioventricular node and ventricles should be normal if the conduction system is intact.
2. Ventricular Pacing
Ventricular pacing is used when the ventricles, the heart’s lower chambers, require direct stimulation. This scenario often occurs in patients with atrioventricular block, where the electrical impulses from the atria do not reach the ventricles.
Mechanism and EKG Changes: In ventricular pacing, the pacemaker delivers impulses directly to the ventricles, typically via a lead in the right ventricle. This direct stimulation causes the ventricles to depolarize in a manner that does not follow the normal conduction pathways, resulting in a widened QRS complex on the EKG. The pacing spike appears just before the QRS complex. Because the impulse starts in the right ventricle, it spreads to the left ventricle in a way that mimics a left bundle branch block (LBBB) pattern.
Clinical Considerations: Diagnosing myocardial infarction (heart attack) in patients with ventricular pacing can be challenging due to these EKG changes. The modified Sgarbossa criteria are used to identify acute myocardial infarction in paced rhythms by assessing specific EKG patterns.
3. Dual Chamber Pacing
Dual chamber pacing involves pacing both the atria and the ventricles, allowing for coordinated contraction and maintaining the natural sequence of heartbeats. This type of pacing is particularly useful for patients who need support in both chambers to optimize cardiac output.
EKG Interpretation: In dual chamber pacing, pacing spikes may appear before both the P-wave and the QRS complex. The pacemaker can sense the heart’s intrinsic activity and only deliver impulses when necessary. If a normal P-wave is detected, the device may inhibit atrial pacing but still pace the ventricles if needed. This flexibility helps maintain a more natural heart rhythm and improves the efficiency of the heart’s pumping action.
4. Biventricular (BiV) Pacing
Biventricular pacing, also known as cardiac resynchronization therapy (CRT), involves simultaneous pacing of both the right and left ventricles. This approach is used in patients with severe heart failure and intraventricular conduction delays that cause the ventricles to contract out of sync, reducing the heart’s efficiency.
Mechanism and EKG Changes: By pacing both ventricles at the same time, biventricular pacing helps synchronize their contractions, improving cardiac output and alleviating heart failure symptoms. On the EKG, this type of pacing may show unique patterns, such as QS or QR complexes in lead I and R or rS patterns in lead V1. These changes reflect the altered depolarization pathways due to the simultaneous stimulation of both ventricles.
Understanding Pacing Spikes on the EKG
Pacing spikes are sharp, vertical lines on the EKG that represent the electrical impulses delivered by the pacemaker. The presence and location of these spikes provide valuable information about which chamber is being paced.
• Atrial Pacing Spike: Appears just before the P-wave, indicating atrial stimulation.
• Ventricular Pacing Spike: Occurs immediately before the QRS complex, indicating ventricular stimulation.
• Dual Chamber Pacing Spikes: Both atrial and ventricular spikes are present, each preceding their respective depolarization waves.
The morphology of the subsequent waves helps assess the effectiveness of the pacing and the heart’s response. Abnormalities in the spikes or the associated waves can signal issues such as lead displacement, battery depletion, or pacemaker malfunction.
Mechanisms of Changes on an EKG in Pacing
Understanding the mechanisms behind the changes seen on the EKG with different pacing modes is crucial for accurate interpretation and patient management.
• Altered Conduction Pathways: Paced electrical impulses may not follow the heart’s normal conduction pathways, leading to changes in the duration and shape of the EKG waves. For example, ventricular pacing causes the impulse to spread through the ventricular muscle rather than the specialized conduction fibers, resulting in a widened QRS complex.
• Bundle Branch Block Patterns: Ventricular pacing, especially from the right ventricle, often produces an EKG pattern similar to a left bundle branch block. Recognizing this pattern is important to avoid misdiagnosing conduction system diseases.
• Assessment of Ischemia: In paced rhythms, traditional criteria for detecting ischemia or infarction may not apply due to the altered EKG patterns. The modified Sgarbossa criteria help identify acute myocardial infarction in these patients by evaluating specific concordant and discordant changes in the ST segments relative to the QRS complexes.
• Pacemaker Malfunction Detection: Abnormalities such as failure to capture (the heart does not respond to the pacemaker’s impulse) or failure to sense (the pacemaker does not detect intrinsic heart activity) can be identified by analyzing the EKG. This analysis is essential for timely intervention and correction of any device-related issues.