Two fundamental mechanisms that enable pacemakers to function effectively are sensing and capture. A thorough understanding of these mechanisms is essential for healthcare professionals to manage patients with pacemakers and accurately interpret EKGs. This article offers an analytical overview of sensing and capture in pacemakers, exploring their functions, underlying mechanisms, and clinical implications.
Sensing in Pacemakers
What does it mean when a pacemaker is sensing?
Sensing refers to a pacemaker’s ability to detect the heart’s intrinsic electrical activity. By monitoring electrical signals through its leads, the pacemaker determines whether it needs to deliver a pacing pulse.
Function of Sensing
The primary function of sensing is to ensure that the pacemaker provides electrical stimulation only when the heart’s natural rhythm is insufficient. If the pacemaker detects an intrinsic heartbeat, it inhibits its own output to prevent unnecessary pacing, thereby conserving battery life and maintaining physiological heart function.
Chambers Involved in Sensing
Sensing can occur in different heart chambers:
• Atrial Sensing: Detects electrical activity in the atria (upper chambers).
• Ventricular Sensing: Detects electrical activity in the ventricles (lower chambers).
• Dual-Chamber Sensing: Monitors both atria and ventricles.
The specific chambers where sensing occurs are indicated by the second letter in the pacemaker’s code. For example, “A” stands for atrial, “V” for ventricular, and “D” for dual.
Purpose of Sensing
Sensing allows the pacemaker to work in harmony with the heart’s natural activity, providing support only when needed. This synchronization helps maintain an appropriate heart rate without interfering unnecessarily with the heart’s own rhythm.
Abnormalities related to sensing
Undersensing
Undersensing happens when the pacemaker fails to detect intrinsic cardiac activity that is present. Consequently, it delivers unnecessary pacing pulses, potentially causing competition between the pacemaker and the heart’s natural rhythm.
Mechanisms of Undersensing
• Inadequate Sensitivity Settings: The pacemaker’s sensitivity may be set too low to detect smaller electrical signals.
• Electrode Dislodgment: The pacing lead may have shifted, reducing its ability to detect cardiac signals.
• Interference: External or internal electrical signals can mask the heart’s activity.
• Lead Damage: Fractured or damaged leads can impair signal detection.
Clinical Implications of Undersensing
Undersensing can lead to inappropriate pacing and arrhythmias. Recognizing undersensing on an EKG allows for timely interventions, such as reprogramming the pacemaker or correcting hardware issues.
Oversensing
Oversensing occurs when the pacemaker detects electrical signals that are not true cardiac activity, leading it to withhold necessary pacing pulses.
Mechanisms of Oversensing
• Electromagnetic Interference: External devices can emit signals that the pacemaker misinterprets.
• Muscle Potentials: Electrical activity from skeletal muscles can be sensed erroneously.
• T-Wave Oversensing: The pacemaker mistakes the T-wave for a QRS complex.
Clinical Implications of Oversensing
Oversensing can result in bradycardia or pauses in the heart rhythm, causing symptoms like dizziness or syncope. Identifying oversensing is crucial for adjusting the pacemaker’s sensitivity or addressing contributing factors.
Capture in Pacemakers
What does it mean for a pacemaker to capture?
Capture refers to the successful depolarization of cardiac tissue following a pacemaker’s electrical impulse, resulting in a heart contraction. On an EKG, capture is indicated by a pacemaker spike immediately followed by the appropriate waveform—a P-wave for atrial capture or a QRS complex for ventricular capture.
Types of Capture
• Atrial Capture: A pacing spike followed by a P-wave indicates the atria have “captured” the pacemaker spike resulting in atrial depolarization.
• Ventricular Capture: A pacing spike followed by a QRS complex indicates the ventricles have “captured” the pacemaker spike resulting in ventricular depolarization.
• Dual Capture: Spikes precede both the P-wave and the QRS complex in dual-chamber pacing.
Abnormalities related to capture
Loss of Capture
Loss of capture occurs when a pacemaker’s electrical impulse fails to depolarize cardiac tissue, evident as a pacing spike not followed by the expected waveform on an EKG.
Mechanisms of Loss of Capture
• Lead Dislodgment: The pacing lead may have moved, losing effective contact with the heart muscle.
• Battery Depletion: Low battery power can reduce the energy of pacing impulses.
• Increased Pacing Threshold: The heart muscle may require a higher energy to depolarize due to factors like ischemia.
• Lead Integrity Issues: Damage to the lead can impair impulse delivery.
Clinical Implications of Loss of Capture
Loss of capture can lead to inadequate heart rates and reduced cardiac output, causing symptoms like fatigue or syncope. It requires immediate evaluation to adjust pacemaker settings or address hardware problems.